Perform Cable IR Drop Calibration with GRL-USB-PD-C2 / C2-EPR

Overview

Using the GRL-USB-PD-C2 / C2-EPR Browser Application, the user can check the impedance level of the cable used for testing. 

Procedure

  1. In the Options tab, select Cable IR Drop Calibration.
  2. Enter a preferred name for the cable.
  3. Select the cable type. 
  4. Click on the Calibrate button to start the IR drop calibration. 
  5. The status will be updated after the calibration is completed.