1. Knowledge Base
  2. Test Solutions
  3. GRL-WP-TPR-C3 Automation Test Browser Application

How to Use API Mode in GRL-WP-TPR-C3 Browser App

Procedure

  1. Open the GRL-WP-TPR-C3 browser application.
  2. Switch from the CTS mode to the API mode as shown below.
  3. After switching to the API mode, you will be able to view only the results panel and all other panels are masked.
  4. Go to "C:\GRL\GRL-WP-TPR-C3\APILibrary" and select GRLC3ApiLibTestingTool as shown below. This will open the "GRLC3ApiLibTestingTool" pop-up window as shown below.

  5. In the "GRLC3ApiLibTestingTool" pop-up window as shown below, click on the Connect button. The "Status", "Version no" and "Serial No" fields will be updated with respect to the GRL-WP-TPR-C3 tester hardware display.
  6. Select the Coil type, click on the Start Capture button and click on the Read Values button. You will be able to view the QI messages, e.g., start time, stop time and CE packets, etc., as shown in the example below.
  7. After test execution has completed, click on the Stop Capture button as shown below.